Complications Associated with Pedicle Screw Placement Using Cortical Bone Trajectory

Yusuf T. Akpolat, Olumide A.Danisa, Wayne K. Cheng


DOI10.4172/2471-8416.100002

Yusuf T. Akpolat, Olumide A.Danisa, Wayne K. Cheng*

Department of Orthopaedic Surgery, Loma Linda University, CANADA

*Corresponding Author:
Wayne K. Cheng
Department of Orthopaedic Surgery
Loma Linda University, CANADA
Tel: (909) 558-6444
Fax: (909) 558-6118
E-mail: md4spine@yahoo.com

Received date: September 21, 2015; Accepted date: September 23, 2015; Published date: September 28, 2015

Citation: Akpolat YT, Danisa OA, Cheng WK. Complications Associated with Pedicle Screw Placement Using Cortical Bone Trajectory. J Clin Exp Orthop. 2015, 1:2. doi: 10.4172/2471-8416.100002

Copyright: © 2015 Cheng WK, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Visit for more related articles at Journal of Clinical & Experimental Orthopaedics

Abstract

Pedicle screw placement usingcortical bone trajectory has become a popular method of fixation especially in minimal invasive spine surgery over the last few years. Complications of this technique have been rarely reported in the literature. A clear understanding of fractures and screw loosening is crucial before embarking on this new surgical technique. Fatigue failure in osteoporotic bone using this technique is alarming and warrants further study to understand its indication.

Introduction

Cortical bone trajectory (CBT) technique for pedicle screw placement in the lumbar spine has described first by Santoni in 2009 [1]. The advantages of using CBT technique involve increased screw purchase within cortical bone and reduced surgical dissection when compared to the traditional pedicle screw (Figure 1) [2-4]. These factors make CBT technique an attractive alternative to standard pedicle screw fixation. This technique is particularly attractive in patients who have compromised bone quality. Several biomechanical studies [1,5-10] were able to demonstrate comparable pull out and toggle strength between CBT and traditional pedicle screw technique. In contrary to these favourable biomechanical results, there is limited data on complications of CBT. In one study Glennie et al. [11] reported screw loosening in 5 of 8 patients. Following are some of the complication we have encountered over the last four years when placing pedicle screws using the CBT.

orthopedics-Traditional-pedicle-screw

Figure 1: Traditional pedicle screw (A) vs. Cortical Bone Trajectory Screw (B).

Pars/pedicle Fractures: In our clinical series, we experienced pars/pedicle fracture with CBT during both intraoperative as well as post-operative period. The fracture started at the entry point, which was medial to the lateral pars; continued through the superior facet then exited out to the lateral aspect of pedicle. We were able to duplicate this identical fracture during our cadaveric study. (Figure 2)

orthopedics-Traditional-pedicle-screw

Figure 2: Figure A showing pars/pedicle fracture on immediate post-op CT scan. Fracture had occurred intra-operatively while using CBT. Screw placed with CBC was removed and replaced with pedicle screw using traditional trajectory. Figure B showing fracture on human cadaveric vertebra. The fracture started at the screw insertion point on the lateral border of the pars, extended in a semi-sagittal oblique plane through the medial aspect of the superior facet, continued through the pedicle, then exited out the lateral aspect of the pedicle.

Screw Loosening: Four out of the initial 26 patients demonstrated radiographic lucency surrounding pedicle screws within 3 month after their initial surgery. This was clearly demonstrated on their CT myelograms, which raised concerns of pseudo arthritis (Figure 3). In addition to our clinical experience, biomechanical studies were performed in our lab to compare fatigue strength of CBT to traditional pedicle screw using osteoporotic cadavers. Side by side comparisons were performed on 12 cadaveric vertebrae. Pedicle screws placed using traditional trajectory outperformed screws placed using cortical bone trajectory based on fatigue testing in osteoporotic vertebrae. CBT screws reached relative motion of 15.2 degree (SD: 5.5), while traditional pedicle screw’s relative motion was 6.9 degree (SD: 4.8) after 100 cycle with 4 Nm load. Pedicle screw placement using cortical bone trajectory has become a popular method of fixation especially in minimal invasivespine surgery over the last few years. Complications of this technique have been rarely reported in the literature. A clear understanding of fractures and screw loosening is crucial before embarking on this new surgical technique. Fatigue failure in osteoporotic bone using this technique is alarming and warrants further study to understand its indication.

orthopedics-Lucency-around-cortical-bone

Figure 3: Lucency around cortical bone trajectory pedicle screws indicating screw loosening.

References

  1. Santoni BG, Hynes RA, McGilvray KC, Rodriguez-Canessa G, Lyons AS, et al.(2009)Cortical bone trajectory for lumbar pedicle screws. Spine J 9: 366-373.
  2. Lee GW, Son JH, Ahn MW, Kim HJ, Yeom JS, et al.(2015)The Comparison of Pedicle Screw and Cortical Screw in Posterior Lumbar Inter-body Fusion: a Prospective Randomized Non-inferiority Trial. Spine J 15: 1519-1526.
  3. Pacione D,Kim I, Wilson TA, Frempong-BoaduA (2015) Cortical screw trajectory for instrumentation and fusion in the setting of osteopathic compression fracture allows for percutaneous kyphoplasty for adjacent level compression fractures. J ClinNeurosci 22: 899-904.
  4. Takata Y, Matsuura T, Higashino K, Sakai T, Mishiro T, et al.(2014) Hybrid technique of cortical bone trajectory and pedicle screwing for minimally invasive spine reconstruction surgery: a technical note. J Med Invest61: 388-392.
  5. Perez-OrriboL,Kalb S, Reyes PM, Chang SW, Crawford NR, et al. (2013) Biomechanics of lumbar cortical screw-rod fixation versus pedicle screw-rod fixation with and without interbody support. Spine (Phila Pa 1976) 38: 635-641.
  6. Baluch DA,Patel AA, Lullo B, Havey RM, Voronov LI, et al. (2014)Effect of Physiologic Loads on Cortical and Traditional Pedicle Screw Fixation. Spine 39: E1297-1302.
  7. Matsukawa K,Yato Y, Kato T, Imabayashi H, Asazuma T, et al. (2014) In vivo analysis of insertional torque during pedicle screwing using cortical bone trajectory technique. Spine (Phila Pa 1976) 39: E240-245.
  8. Inceoglu S, Montgomery WH Jr, St Clair S, McLain RF (2011) Pedicle screw insertion angle and pull out strength: comparison of 2 proposed strategies. J Neurosurg Spine 14: 670-676.
  9. Calvert GC, Lawrence BD, Abtahi AM, Bachus KN, Brodke DS, et al. (2015)Cortical screws used to rescue failed lumbar pedicle screw construct: a biomechanical analysis. J Neurosurg Spine 22: 166-172.
  10. WrayS,Mimran R, Vadapalli S, Shetye SS, McGilvray KC, et al. (2015) Pedicle screw placement in the lumbar spine: effect of trajectory and screw design on acute biomechanical purchase. J Neurosurg Spine 22: 503-510.
  11. Andrew G R,Dea N, Kwon BK, Street JT (2015) Early clinical results with cortically based pedicle screw trajectory for fusion of the degenerative lumbar spine. J ClinNeurosci22: 972-975.
open access journals, open access scientific research publisher, open access publisher
Select your language of interest to view the total content in your interested language

Viewing options

Flyer image

Share This Article